Aris Widodo, AW and endarko, E (2018) Experimental study of one step linear Gauss- Newton algorithm for improving the quality of image reconstruction in high-speed Electrical Impedance Tomography (EIT). Journal of Physics: Conference Series, 1120 (1). pp. 1-8. ISSN 1742-6588
Text
Widodo_2018_J._Phys.__Conf._Ser._1120_012067.pdf Download (1MB) |
Abstraksi
Abstract. One step linear Gauss-Newton was well known as one of image reconstruction algorithm in Electrical Impedance Tomography (EIT). In the present study, we proposed the mathematical model of one step linear Gauss-Newton for processing the experimental data for achieving the best image reconstruction. Two ways were conducted to solve this study. The first one, we build EIT hardware that mainly controlled by Single Board Computer (SBC) to produce a constant current at 1 mA 20 kHz that will be injected to the 16 electrodes of the practical phantom with speed of 1 ms/data for acquiring the data of voltage boundary from that practical phantom when using aluminium as anomaly. The second one, we processed the data of voltage boundary using one-step linear Gauss-Newton to achieve the best image reconstruction with modified the exponent (p) of Regularization matrix that varied at 0 1 by the increment of 0.1 and hyperparameter value (λ). The result showed that the best image reconstruction obtained at p = 0.7 and λ = 8 for this algorithm.
Item Type: | Article |
---|---|
Subjects: | Q Science > QC Physics |
Divisions: | Fakultas Sains, Teknologi dan Pendidikan |
Depositing User: | Dosen Aris Widodo |
Date Deposited: | 09 Feb 2023 06:43 |
Last Modified: | 09 Feb 2023 06:43 |
URI: | http://repository.umla.ac.id/id/eprint/2531 |
Actions (login required)
View Item |